7.9 Technical specifications

Overview
Screen technologyIPS
Active screen size (diagonal)853.44 mm (33.6")
Active screen size (H x V)708.1 x 472.1 mm (27.8 x 18.6")
Aspect ratio (H:V)3:2
Resolution12MP (4200 x 2800 pixels)
Pixel pitch0.1686 mm
Color imagingYes
Gray imagingYes
Bit depth30 bit
Viewing angle (H, V)178°
Optical glassYes
Uniformity correctionPPU
SteadyColor CalibrationYes
I-LuminateYes
Ambient Light Compensation (ALC)Yes, reading room selection
Ambient light sensorYes
Backlight Output Stabilization (BLOS)Yes
Front sensorYes, I-Guard
Maximum luminance2100 cd/m² (PPU on)
DICOM calibrated luminance1000 cd/m²
Contrast ratio (panel typical)1200:1 (PPU on)
Response time ((Tr + Tf)/2) (typical)16.5 ms
Housing colorRAL 9004 / RAL 9003
Video input signalsDisplayPort 1.2
USB ports

1x USB 2.0 upstream (endpoint)

3x USB 2.0 downstream

Power rating24 VDC, 16.25 A; 5 VDC, 0.1 A
Power requirements

This device shall only be powered by the medical approved power supply of Efore (Roal Electronics), type RHPS390.

Ratings marked on the medical approved power supply:

  • Input rating: 100-240 VAC, 5.5 A, 50/60 Hz
  • Output rating: 24 VDC, 16.25 A; 5 VDC, 0.1 A
Power consumption

190 W (nominal)

< 0.5 W (hibernate)

Dimensions with stand (W x H x D)795 x 610 x 300 mm (lowest position)
Dimensions w/o stand (W x H x D)795 x 572 x 131 mm
Dimensions packaged (W x H x D)960 x 715 x 395 mm
Net weight with stand33 kg
Net weight w/o stand23 kg
Net weight packaged42 kg
Tilt-5° / +30°
Swivel-23° / +23°
PivotN/A
Height adjustment range95 mm
Mounting standardVESA (200 x 100 mm & 100 x 100 mm)
Screen protectionProtective, non-reflective glass cover
Recommended modalitiesAll digital images, including digital mammography and breast tomosynthesis
Certifications

FDA 510(K) K151505 for General Radiology

CE0123 (Medical Device)

CCC (China), KC (Korea), Inmetro (Brazil), BIS (India), NOM (Mexico), EAC (Russia, Kazakhstan, Belarus, Armenia and Kyrgyzstan)

Safety specific:

IEC 60950-1:2005 + A1:2009

EN 60950-1:2006 + A1:2010 + A11:2009 + A12:2011 + A2:2013

IEC 60601-1:2005 + A1:2012

EN 60601-1:2006 + A1:2013 + A12:2014

ANSI/AAMI ES 60601-1:2005 + R1:2012

CAN/CSA C22.2 No. 60601-1:14

EMI specific:

IEC 60601-1-2:2014 (ed4)

EN 60601-1-2:2015 (ed4)

FCC part 15 Class B

ICES-001 Level B

VCCI (Japan)

Environmental:

China Energy Label, EU RoHS, China RoHS, REACH, Canada Health, WEEE, Packaging Directive

Supplied accessories

User Guide

Quick Installation Sheet

System Sheet

Video cables

Main cables

USB 2.0 cable

External power supply

Film clip

Barco Touchpad

Optional accessoriesNone
QA softwareQAWeb
Warranty5 years, including 40000 hours backlight warranty
Operating temperature0°C to +35°C (+20°C to +30°C within spec)
Storage temperature-20°C to +60°C
Operating humidity20% - 85% (non-condensing)
Storage humidity20% - 85% (non-condensing)
Operating pressure70 kPa minimum
Storage pressure50 to 106 kPa