Electrostatic discharge (ESD) IEC 61000-4-2 | ± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air | ± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air | Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30% |
Electrical fast transient/burst IEC 61000-4-4 | ± 2 kV for power supply lines ± 1 kV for input/ output lines 100 kHz repetition frequency | ± 2 kV for power supply lines ± 1 kV for input/ output lines 100 kHz repetition frequency | Mains power quality should be that of a typical commercial or hospital environment |
Surge IEC61000-4-5 | Line to line: ± 0.5 kV, ± 1 kV Line to ground: ± 0.5 kV, ± 1 kV, ± 2 kV | Line to line: ± 0.5 kV, ± 1 kV Line to ground: ± 0.5 kV, ± 1 kV, ± 2 kV | Mains power quality should be that of a typical commercial or hospital environment |
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 | 0% residual voltage for 0.5 period at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% residual voltage for 1 period at 0° 70% residual voltage for 25 periods at 0° Voltage interruptions: 0% residual voltage for 250 periods at 0° | 0% residual voltage for 0.5 period at 0°, 45°, 90°, 135°, 180°, 225°, 270° and 315° 0% residual voltage for 1 period at 0° 70% residual voltage for 25 periods at 0° Voltage interruptions: 0% residual voltage for 250 periods at 0° | Mains power quality should be that of a typical commercial or hospital environment. If the user of the MDNC-3421 requires continued operation during power mains interruptions, it is recommended that the MDNC-3421 be powered from an uninterruptible power supply or a battery |