About DICOM status report
- DICOM Compliance status: Shows if the current DICOM curve is compliant or not.
- Maximum error: Shows the maximum error of the current DICOM curve. This is the deviation compared to a perfect DICOM.
- Luminance: Shows the measured luminance.
- Black luminance: Shows the measured black luminance.
- Time elapsed since latest compliance check: Shows the backlight runtime since last compliance check.
DICOM Calibration Status
- Not executed: No other information is visible.
- Calibration executed: When the calibration is executed, the time elapsed since latest calibration is shown.
- Time elapsed since latest calibration: Shows the backlight runtime since last calibration check.
- DICOM compliance check: Measures the DICOM curve of your display in different steps.
- DICOM calibration: Adds a correction to the current DICOM curve to approach the perfect DICOM curve as well as possible.
- DICOM Error threshold: Sets the error threshold. This is the maximum error allowed before a DICOM calibration is required.
- Ambient Light Compensation: Disables/Enables the ambient light compensation function.
- Reading Room: Selects a reading room.
- Reset DICOM Calibration: Restores the original (factory default) DICOM curve.