Troubleshooting SteadyColor™ Compliance Test Failures for MDNC-6121 Displays in MediCal QAWeb

Article number: [5235] - Legacy code: [6425]

Applicable to

Please find more info about Barco SteadyColor™ technology here and see how this technology works and which clinical impact it has.

Your MDNC-6121 display has been calibrated and still, it is failing the SteadyColor™ compliance test. There may be several reasons for that.

  1. You are using an incompatible display controller
  2. Your display controller drivers are outdated
  3. You are not using a compatible QAWeb Agent version

1 You are using an incompatible display controller

SteadyColor™ calibration for the MDNC-6121 is only supported on MXRT-5600 and MXRT-7600 display controllers. If you are using older MXRT display controllers or 3rd party display controllers, this SteadyColor™ calibration is not available.

You can either use a compatible display controller or disable the Color Compliance Test in MediCal QAWeb. See the following article: [kb7974]

2 Your display controller drivers are outdated

Along with the requirements of the display controller, a minimum MXRT driver version of 10.143. Barco 2.0 is required. Consult the following article to find out where you can find the latest MXRT drivers: [KB4010].

3 You are not using a compatible QAWeb Agent version

Steadycolor™ calibration for the MDNC-6121 is only supported by the QAWeb Agent from a certain version onwards. When using older versions, MediCal QAWeb will only perform a regular DICOM calibration and therefore your display will not receive the appropriate color calibration which is recommended for this display.

Depending on the type of MDNC-6121 display, you will need the following QAWeb Agent versions:

  • MDNC-6121 (part number K9300360A): QAWeb Agent 1.13.10 or higher
  • MDNC-6121 NC (part number K9300361A): QAWeb Agent 1.13.17 or higher
Note: The NC model (part number K9300361A) is the model without a protective cover and without glass in front of the LCD panel.

Further reading:

  • [KB6427] How to check the currently installed version of MediCal QAWeb Agent
  • [KB6430] Where to find MediCal QAWeb Agent and Relay installation files

Related files

Properties

Last updated Jul 23, 2024

No solution found?

Phone support

Our helpdesk provides you with prompt phone support. A team of experienced support engineers is at your service for any professional assistance.